EasyManua.ls Logo

ABB AquaMaster4 - Page 55

ABB AquaMaster4
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AQUAMASTER4 | ELECTROMAGNETIC FLOWMETER TRANSMITTER | OI/FET400-EN REV. D 55
Cellular 4G settings* Description Description
Network operator Operator name of the network that the device registered to Read-only
Signal quality Cellular signal quality in percentage Read-only
Local battery voltage Current voltage of cellular interface buffer battery Read-only
Access technology Current selected cellular access technology. Options: Auto | UTRAN | UTRAN w/ HSDPA | UTRAN w/
HSUPA | UTRAN w/ HSDPA n HSUPA | E-UTRAN | CDMA
Read-only
Roaming indicator Indicator for roaming status. Options: Off/On Read-only
Data report timebase Time base for data report with one day in seconds. Range: 0 ~ 86399 seconds (i.e. within 1 day) Read-only
Data report interval Time interval for data report Read-only
Data report mode The type of data report to be sent. Options: Summary report/Detail report Read-only
File format Current file format for cellular information exchange. Options: JSON | CSV Read-only
FTP server host name Remote FTP Server host name in IP address or domain name format. Read-only
FTP server port Remove FTP server port number Read-only
FTP user name Current user name for accessing remote FTP Server Read-only
FTP working directory Current working directory on remote FTP server Read-only
TLS on off TLS is enabled or not Read-only
TLS auth mode Current TLS authentication method. Options: None/Verify server cert only/Mutual authentication Read-only
DNS primary server address IP address of primary DNS server Read-only
DNS secondary server address IP address of secondary DNS server Read-only
NTP server host name Remote NTP Server host name in IP address or domain name format Read-only
NTP server port Remote NTP Server port number Read-only
Trigger data report Manually trigger a data report Editable
Last sent report time Display the last data report sent time Read-only
Cellular NB settings* Description Note
Signal quality Cellular signal quality in percentage Read-only
Local battery voltage Current voltage of cellular interface buffer battery Read-only
Data report timebase Time base for data report with in one day in seconds. Range: 0 ~ 86399 seconds (i.e. within 1 day) Read-only
Data report interval Time interval for data report. Options: 1 hour | 6 hours | 12 hours | 24 hours Read-only
Server IP IP address of IoT Platform to connect Read-only
Server port Port number of IoT Platform to connect. Read-only
Band Current cellular band configuration. Options: Band 1 | Band 3 | Band 5 | Band 8 | Band 20 | Band 28 Read-only
Power class Current cellular power class configuration. Options: 23dBm | 20dBm | 14dBm Read-only
Trigger data report Manually trigger a data report Editable
Firmware Information Description Note
Flowmeter Firmware Part Code** 3KXF208402U0113
Flowmeter Firmware Version** The version of firmware installed on the transmitter
Flowmeter Firmware Date** The firmware date on the transmitter
Bootloader Firmware Version** ID of the bootloader firmware
UAM Firmware Version** ID of the update application manager firmware
MCU0 Application Version* The version of firmware installed on the transmitter Read-only
MCU0 Boot0 Version* The version of firmware installed on the transmitter Read-only
MCU0 Boot1 Version* The version of firmware installed on the transmitter Read-only
MCU1 Application Version* The version of firmware installed on the transmitter Read-only
MCU1 Boot0 Version* The version of firmware installed on the transmitter Read-only
MCU1 Boot1 Version* The version of firmware installed on the transmitter Read-only
Range Error List
System Error Message
Reboot System Use to reboot the system Editable
* Only available with FEX45X transmitter.
** Only available with FEX41X/43X transmitter.

Table of Contents