Connecting the CoreSense M10 47
Connecting the internal communication cable
The sensor head and analytical unit communicate via a proprietary communication cable.
To connect the communication cable:
1 Of the cables coming from the sensor head conduit, identify the one with the orange and blue
internal wires.
—
Figure 25 Communication cable configuration
TX+
TX–
RX+
RX–
GND
2 Connect each of the five wires in the appropriate socket on the HU485 connector (see below).
—
Figure 26 Analytical unit connection board
HU485 connector Sensor head DC power connector
NOTICE
For more information on the various communication cables used, see Table 4 on
page 24 and “Connector definitions” on page 89.
For more information on configuring the communication protocols, refer to the
CoreSense M10 Monitoring System User Guide.