FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D  113 
 
 
Error no. / Range  Text on the LCD display  Cause  Remedy 
S041.034 / electronics  DC feedback  regulation. 
Check conditions of application. 
Call service. 
Multi-phase measuring media that produce a very 
high level of noise. 
Stones or solids that produce a very high level of 
noise. 
Galvanic voltages at the measuring electrodes. 
Conductivity of the measuring medium is not 
evenly distributed (e.g. directly after the injection 
points). 
Please contact the service department. 
S040.031 / electronics  Coil Inductance alarm. 
Call service. 
Coil inductance changed, coil damaged, coil 
insulation damaged, external magnetic fields. 
Please contact the service department. 
 
Maintenance 
 
Error no. / Range  Text on the LCD display  Cause  Remedy 
M038.009 / electronics  Sensor memory defective. 
Mem. or connect. defective. 
Replace memory. 
Defective NV memory module. 
NV memory module not inserted. 
Check if NV memory module is inserted. 
Please contact the service department. 
M037.014 / Electronics  NV chips defect on 
Motherboard. 
Defective MB. Replace MB. Call 
Service. 
Defective NV memory.  Replace motherboard. 
Please contact the service department. 
M032.022 / Config.  Curr.Out 31 / 32 not calibrated. 
Call Service. 
Current output 31 / 32, Uco not calibrated.  Please contact the service department. 
 
M031.023 / Config.  Curr.Out V1 / V2 not calibrated. 
Replace Current Option Card. 
Call Service. 
Current output V1 / V2 not calibrated.  Please contact the service department. 
 
M030.024 / Config.  Curr.Out 31 / 32 not calibrated. 
Call Service. 
Current output V3 / V4 not calibrated.  Please contact the service department. 
 
M028.007 / Config.  Display value is < 1600 h at 
Qmax. 
Change mass Unit or vol. Unit 
for Totalizer. 
Counter unit too small.   Change mass or volume totalizer unit. 
 
M020.027 / Electronics unit  Communicat. Card not 
responding 
Fieldbus plug-in is not reacting. 
•  Plug-in card is defective. 
Please contact the service department. 
 
M026.004 / Config.  Maintenance interval is reached. 
Perform maintenance. 
Set ‘Preset Maint. cycle’ to zero to disable the 
maintenance timer. 
Perform maintenance work. 
Start new cycle. 
M024.008 / Config.  Device not  calibrated. 
Call Service. 
Device is not calibrated.  Please contact the service department.