FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D  131 
 
13 Specification 
 
Note 
The device data sheet is available in the ABB download area at 
www.abb.com/flow. 
 
 
Permitted pipe vibration 
 
In accordance with EN 60068-2-6 
Valid for sensors in remote mount and integral mount design. 
•  Maximum deflection: 0.15 mm (0.006 in) in the frequency 
range of 10 to 58 Hz 
•  Maximum acceleration: 2 g in the frequency range of 
58 to 150 Hz 
 
 
ProcessMaster – Temperature data 
 
The temperature range offered by the device is dependent on a 
number of different factors. 
These factors include the measuring medium temperature 
T
medium
, the ambient temperature T
amb
, operating pressure 
P
medium
, liner material and the approval for explosion protection. 
 
 
Storage temperature range 
−40 to 70 °C (−40 to 158 °F) 
 
 
 
 
 
 
Maximum permissible cleaning temperature 
 
CIP media  Liner
Cleaning temperature 
Steam PTFE, PFA
150 °C (302 °F) 
Cleaning fluid  PTFE, PFA
140 °C (284 °F) 
 
•  The maximum cleaning temperature specified applies to a 
maximum ambient temperature of 25 °C (77 °F). 
  If the ambient temperature up-scales > 25 °C (> 77 °F), then 
the temperature difference to the current temperature must 
be subtracted from the max. cleaning temperature. 
•  The specified cleaning temperature may have an effect for a 
maximum of 60 minutes. 
 
 
Change from two to one  column