Monitoring the grounding 
The function checks for noise in the measuring signal and the 
electrical grounding of the device. While the check is in progress, 
no flow measurement can take place. 
The noise / grounding check is started manually and delivers a 
‘successful / failed’ result. 
The measurements (Power Spectrum, Amplitude 1 to 4 and 
Frequency 1 to 4) will help if the noise / grounding check fails. 
 
Requirements for use: 
•  The sensor must be filled completely with measuring 
medium. 
•  There is no flow through the sensor (close all valves, shut-
off devices etc.) 
•  The sensor must be grounded (see Sensor grounding on 
page 29). 
•  There may not be any deposits on the measuring 
electrodes. 
 
 
 
Menu / parameter  Description 
Diagnostics / ...Diagnosis Control / ...Noise Check 
Start Noise Check  Start of test 
Result Noise Check  Test result 
Power Spectrum  Current power spectrum. 
Amplitude 1 Value ... 4  Display of the four strongest amplitudes of the 
frequency spectrum in μV with the associated 
frequency. 
Frequency 1 ... 4 
 
 
 
 
 
 
Verification 
 
 
Fingerprint database 
The sensor and transmitter fingerprint stored in the 
SensorMemory allows you to compare the state of the device at 
the time of manufacture at the factory with the current state of 
the device at the customer site. 
 
The check is started manually and returns a ‘successful / failed’ 
result. 
 
If the verification is unsuccessful, troubleshooting information is 
shown on the display (parameter ‘Rslt FP Verification‘). 
 
A software tool (ABB Ability SRV500) is available for 
documentation and trend analysis. 
 
 
Setup 
 
Menu / parameter  Description 
Diagnostics / ...Diagnosis Control / ...Fingerprints 
Tx Factory CMR, 1m/s, 
10m/s 
Display of transmitter fingerprint (factory 
fingerprint) 
Se Factory Coil Ind.  Display coil impedance fingerprint 
Se Factory Imp. E1  Display electrode impedance fingerprint E1-GND, 
E2-GND 
Se Factory Imp.E2 
Strt FP Verification  Start of test 
Rslt FP Verification  Test result 
Tx Customer CMR, 1m/s, 
10m/s 
Display of transmitter fingerprint (customer 
fingerprint) 
Se Customer Coil Ind  Display coil impedance fingerprint 
Se Customer Imp. E1  Display electrode impedance fingerprint E1-GND, 
E2-GND 
Se Customer Imp. E2