SwirlMaster FSS430, FSS450 VortexMaster FSV430, FSV450 | CI/FSS/FSV430/450-EN Rev. D 29
6.3 Checking and configuring the basic settings
The device can be factory parameterized to customer
specifications upon request. If no customer information is
available, the device is delivered with factory settings.
Parameter Factory setting
Active Mode Liquid Volume
Output Value Flow rate
DO Function No function
Q
max
Actual value set to Q
max
DN.
Depending on the nominal
diameter of the flowmeter.
Unit Q m
3
/h
Analog In Value No function
Hart In Value No function
Low Flow Cutoff 4 %
Iout at Alarm Low Alarm Value
Low Alarm Value 3.55 mA
High Alarm Value 22 mA
Specific operating conditions
If the manufacturer of the device has not stated the type of
protection on the name plate, the user must state the type of
protection used on the name plate during installation.
The painted surface of the device can potentially store
electrostatic charges and represent a potential ignition source
during applications with low relative humidity. Therefore, the
relative humidity must be below 30 % and the painted surface
must be free of surface contaminants such as dirt, dust and
oil. Guidelines on protection against the risk of ignition due to
electrostatic discharge are presented in EN TR50404 and IEC
TR60079-32 (in preparation). The painted surface must be
cleaned using only a moist cloth.
6.3.1 Parameterization via the "Easy Setup" menu
function
Setting of the most current parameters is summarized in the
"Easy Setup" menu. This menu provides the fastest way to
configure the device.
NOTE
The LCD display is provided with capacitive control buttons.
These enable you to control the device through the closed
housing cover.
The following section describes parameterization via the "Easy
Setup" menu function.
QNP 0.00 m3/h
QD 000000 %
QD 000000 %
1. Use to switch to the configuration level.
ccess Level
Read Only
Standard
Service
Back Select
2. Use / to select "Standard".
3. Confirm the selection with .
Enter Password
**********
RSTUVWXYZ 12345
Next OK
4. Use to confirm the password. A password is not
available as factory default; you can continue without
entering a password.
Menu
Easy Setup
Exit Select
5. Use / to select "Easy Setup".
6. Confirm the selection with .