ENGLISH
PSPIX² • User manual
21
WARNING:
Risk of electrocution and/or malfunction of the PSPIX system.
- The PSPIX system must only be connected to a compliant wall plug.
- Before connecting the PSPIX system, check that the mains voltage and frequency indicated on the
power supply correspond to those of mains supply.
- Before switching on, ensure that the neither PSPIX system nor the power supply cable are damaged.
Damaged cables and connectors must be replaced immediately.
- Never touch the patient and non-protected connectors on the PSPIX system at the same time.
- Place the PSPIX system so that you can easily access the electrical socket to disconnect it.
- Where you are using multiple adapters, the requirements of IEC 60601-1 must be observed. Do not
place the multiple adapters on the ground. Other systems must not be connected to the same multiple
adapter.
The PSPIX system allows up to ten computers to be connected.
- Connect the PSPIX system with the supplied Cat5 Ethernet Cable:
• either directly to a computer, when only one workstation is being connected
• or to the existing network when either one or more computers are being connected.
- Ensure that the computer connected to the PSPIX system is connected to a correctly earthed and switched-
on mains socket.
CAUTION:
Risk of electrocution. The computer connected to the PSPIX system must not be used near the patient,
in compliance with IEC standard 60601-1. The minimum horizontal distance between the patient and the
computer must be 1.5 m. The minimum vertical distance between the patient and the computer must
be 2.5 m.
- Once the PSPIX system is connected to the computer or network, install the PSPIX acquisition module on
the computer. This is on the CD-ROM supplied with the system.
-Insert the CD-ROM supplied with the PSPIX system into the computer drive to install the acquisition module.
NOTE:
The SOPRO IMAGING software is also available on the CD-ROM supplied with the PSPIX system . The PSPIX
acquisition module is directly integrated into this software. If you wish to use SOPRO IMAGING, please
install it and follow the instructions displayed on the screen up until the end of the procedure.
If not please refer to the user manual for the imaging software being used.
-Follow the instructions displayed on the screen up until the end of the procedure.
- Repeat this installation process for all computers you wish to connect the PSPIX system to.