7451A/7461A/7461P Digital Multimeter Operation Manual
6.7.3 ADC Command Reference
6-34
System Calibration CAL0 Sets the calibration mode to OFF.
(Writes the calibration factor when leaving the calibration
mode.)
CAL1 Sets the calibration mode to ON.
CAL? Reply: CAL0 or CAL1
XOUT Cancels (sets to OFF) the calibration mode.
(Does not write the calibration factor.)
PCn Enters the STD value. (Displayed count value)
n:0 to 999999
XDTn Enters the STD value. (Displayed value)
n: Displayed value on STD
XTERM? Performs the zero calibration at the rear input terminal.
Reply: 0 (Normal) or 1 (Abnormal)
CMNT
"str"
Sets the calibration information.
Enclose the character strings by using (") or (').
Up to 50 characters (one byte alphanumeric characters) can be
stored.
CMNT?
Reply: CMNT"xxxx xxx"
Setting parameter *SAVn n:0 to 3
Saves the setting parameters in area [n] of the non-volatile
memory.
SINI Sets the values when shipping to all area from [0] to [3].
*RCLn n:0 to 3
Loads the setting parameters from area [n] of the non-volatile
memory.
RINI Loads the values when shipping as the setting parameters.
Error log reading ERR? Reads the error details.
Reply: ddd,"xxxxxxxxxx"
• Up to 20 error queues can be stored. The error is output
according to the FIFO system.
• If more than 20 errors occur, the last error is replaced with -
350, "Queue overflow". Any errors after this are not saved.
• If no error occurs, +000, "No error" is returned.
Input terminal IN0 Front
IN1 Rear
IN? Reply: IN0 or IN1
Item Command Description Initial value
Error character strings
(Up to 80 characters)
Error code