7451A/7461A/7461P Digital Multimeter Operation Manual
6.7.4 SCPI Command Reference
6-37
6.7.4.2 Remote Command
1. Measurement command
Function description SCPI command Parameter Query reply
Setting the DCV measurement
function and sending the read out
request at the same time
:MEASure? Refer to Section 6.7.2,
“Output Data Format.”
1Performing the measurement
once and reading the measure-
ment value
:MEASure:<Function>? Refer to Section 6.7.2,
“Output Data Format.”
<Function> can be selected from the followings:
DCV VOLTage:DC
ACV VOLTage:AC
2W RESistance
4W FRESistance
DCI CURRent:DC
ACI CURRent:AC
ACV(AC+DC) VOLTage:ACDC
ACI(AC+DC) CURRent:ACDC
Diode DIODe
LP-2W RESistance:LPOWer
LP-4W FRESistance:LPOWer
Continuity CONTinuity
Freq FREQuency
TEMP *7461P only TEMPerature
DCV-Bch HI-COM VOLTage:BDC:BCOM
DCV-Bch HI-LO VOLTage:BDC:B
DCV-Bch RATIO VOLTage:BDC:RATio
Presetting the specified function :CONFigure:<Function>
Reading the measurement function
setting
:CONFigure? "<func>,<real>"
(String, Range)
Changing the measurement func-
tion
DCV [:SENSe]:FUNCtion "VOLTage:DC" "VOLT:DC"
ACV "VOLTage:AC" "VOLT:AC"
2W "RESistance" "RES"
4W "FRESistance" "FRES"
DCI "CURRent:DC" "CURR:DC"
ACI "CURRent:AC" "CURR:AC"
ACV(AC+DC) "VOLTage:ACDC" "VOLT:ACDC"
ACI(AC+DC) "CURRent:ACDC" "CURR:ACDC"
Diode "DIODe" "DIOD"