7451A/7461A/7461P Digital Multimeter Operation Manual
6.7.4 SCPI Command Reference
6-38
LP-2W "RESistance:LPOWer" "RES:LPOW"
LP-4W "FRESistance:
LPOWer"
"FRES:LPOW"
Continuity "CONTinuity" "CONT"
Freq "FREQuency" "FREQ"
TEMP *7461P only "TEMPerature" "TEMP"
DCV-Bch HI-COM "VOLTage:BDC:
BCOM"
"VOLT:BDC:BCOM"
DCV-Bch HI-LO "VOLTage:BDC:B" "VOLT:BDC:B"
DCV-Bch RATIO "VOLTage:BDC:
RATio"
"VOLT:BDC:RAT"
Auto-range setting
The setting to Continuity, Diode,
and DCV-Bch is disabled.
[:SENSe]:<Function>:RANGe:AUTO {ON,1|OFF,0} {1|0}
Temperature unit setting
*7461P only
:UNIT:TEMPerature {C|F|K} {C|F|K}
Temperature measurement method
setting
*7461P only
[:SENSe]:TEMPerature:TRANsducer:TYPE {RTD|TRTD|FRTD}
RTD:2W
TRTD:3W
FRTD:4W
{RTD|TRTD|FRTD}
Temperature sensor standard set-
ting
*7461P only
[:SENSe]:TEMPerature:TRANsducer:STANdard {PT100|JPT100} {PT100|JPT100}
Measurement range setting
DCV [:SENSe]:<Function>:RANGe 0 Setting value
Maximum value in the
measurement range
(750 V in the fre-
quency measurement)
<real_6> (7461A/61P)
<real_5> (7451A)
ACV, ACV(AC+DC), Freq
2W, 4W
DCI, ACI, ACI(AC+DC)
LP-2W, LP-4W
The setting to Continuity,
Diode, and DCV-Bch is dis-
abled.
Sampling rate setting [:SENSe]:<Function>:SRATe {FREE|SFASt|FAST|
MED|SLOW|SSLow}
{FREE|SFAS|FAST|
MED|SLOW|SSL} *1
Display digit setting [:SENSe]:<Function>:DIGits *3 [3|6|6] (7461A/61P)
[3|5|5] (7451A)
<int_1>
Integration time setting (PLC)
The setting is disabled in the fre-
quency measurement.
[:SENSe]:<Function>:NPLCycles [0.02|10|100] *2 <real_6>
Integration time setting (second)
The setting is disabled in the fre-
quency measurement.
[:SENSe]:<Function>:APERture [0.00001|10]
(7461A/61P)
[0.0001|10]
(7451A)
<real_6>
*1: ‘FREE’ is returned when the sampling interval or integration time is set.
*2: When the integration time is outside this range, it is set in seconds. (Refer to Section 5.7.1, “Sampling Interval (SI) and Integration Time
(IT) Settings”.)
*3: The setting is applied to all functions.
Function description SCPI command Parameter Query reply