7451A/7461A/7461P Digital Multimeter Operation Manual
7.1.10 Measurement Time and Display Digits
7-18
Integration Time (IT) and Sampling Interval (SI)
Integration Time and Display Digits
7461A/61P 7451A
Integration time (IT) Integration time (IT)
Setting range Resolution Setting range Resolution
10 s to 199 s1 s100 s to 199 s1 s
200 s to 9.99 ms 10 s200 s to 9.99 ms 10 s
10 ms to 5999.9 ms 100 s 10 ms to 5999.9 ms 100 s
6 s to 10 s 200 s 6 s to 10 s 200 s
0.02 PLC to 100 PLC 0.01 PLC 0.02 PLC to 100 PLC 0.01 PLC
Sampling interval (SI) Sampling interval (SI)
Setting range Setting range
50 s to 3600 s 200 s to 3600 s
Function
DCV/DCI
DCV-Bch
2W/4W
LP-2W/4W
ACV/ACI
Diode and Continuity
ACV(AC+DC)
ACI(AC+DC)
Integration time (IT)
7461A/61P 7451A 7461A/61P 7451A
10 s (IT) 200 s 3 1/2-digit display 3 1/2-digit display 3 1/2-digit display
200 s < (IT) 2 ms 4 1/2-digit display 4 1/2-digit display 3 1/2-digit display
2 ms < (IT) 15 ms 5 1/2-digit display 5 1/2-digit display 4 1/2-digit display
15 ms < (IT) 10 s 6 1/2-digit display 5 1/2-digit display 4 1/2-digit display