26
28 30
0
Figure 8 12 Step Results
1. If necessary, press the navigation keys Up Down until the source area is active.
2. Press Step to show the step setup interface, and configure these parameters as follows:
Start Value (in units).
End Value (in units).
Step Mode: Eng.Units, Scale or Step Points.
Step Size (Scale or Count, according to the Selection of Step Mode).
Trans Function: Linear or Square Root.
3. To complete the step settings, press Enter or Done , then the calibrator will Return to the Basic Mode and show the
main screen.
4. Press Manual to start manual stepping. The softkey label will changes to Next.
5. To advance to the next value press Enter or Next.
6. If you want to exit manual stepping, press Esc or Exit.