When the measurement is active in HART communication mode, press Setup to show the HART setup interface.The calibrator
provides five setup functions (refer to table 11-1).
The setting range cannot exceed the transmitter range
upper limit
Float type the value cannot be less than 0
Can be changed into linear transfer or rooting transfer
Read Write
Read Write
Read Write
Damping time
Primary variable range upper limit
Transfer function
N A
N A
N A
Read only
Read only
Read only
Sensor range lower limit
Sensor S N
Sensor range upper limit
Setup
Basic
Sensor
Can be changed into any units supported by transmitter
The setting range cannot be less than transmitter range
lower limit
N A
Read Write
Read Write
Read only
Primary variable range units
English input is available no more than 8 characters
Read WriteDevice Label
Primary variable range lower limit
Sensor minimum span
Attributes DescriptionVariables