44
10 Task and AsCal
AsCal is a typical calibration process including as-found and as-left, and Task is the calibration procedure of a DUT (Device
Under Test).
The ADT22XA calibrators can perform AsCal, calculate errors (Zero Error and Span Error), display results in the memory and
highlight the out-of-tolerance points.
You can download tasks and upload the results with special software. The software can generate reports and certificates for
the tasks.
AsCal will make your calibration faster and easier
The As Found calibration documents the state of the instrument before carrying out any adjustments. The calibration shows
the instrument's level of drift during calibration.
After routine maintenance and deficiencies detected during calibration are addressed an “as-left” calibration is performed.