27
In the Basic Mode, volts, millivolts, frequency, resistance and current can
be reset by pressing the Reset softkey, which will set the source value to
zero.
In Basic Mode, the calibrator supports stepping the source value manually
or automatically. First you will need to configure the step settings. Each
source item (except pulse and pressure) has its own step settings. When
the source is active press Step to view the step setup interface. Figure 8.11
and Figure 8.12 are examples of a step configuration and results.
The start value cannot be the same as the end value, and the values cannot
exceed the output range.
Stepping supports three modes: engineering units, scale and step points.
The run mode and step time are relevant only for auto step. The range of
the step time is 1 to 3600 seconds.
Figure 8 11 Stepping Configuration