40
The calibrator uses the first and last deviation function for the calculation for SPRT If you use other temperature ranges
b
i
and c
i
On the RTDs Management page, you can delete an existing RTD by pressing Delete and re-edit it by pressing Enter
or Edit.
When the custom RTD is added to the RTDs Library successfully, your calibrator is now ready to make measurements
using your custom RTD.
Temperature range(°C)
0 to 29 7646 TPW Ga TP
0 to 156 5985 TPW Ga TP In FP
0 to 231 928 TPW In FP Sn FP
0 to 419 527 TPW Sn FP Zn FP
Required Fixed Points Deviation Function
-189 3442 to 0 01
Ar TP Hg TP TPW
0 to 660 323
TPW Sn FP Zn FP Al FP