46
According to different instruments, there are five types of task: RTDs (Resistance Temperature Detector), TC (Thermocouples),
To start your task management press Task in Basic Mode. Select Task List in Task Management page and press Enter r
Select.On the Task List page, you can create a new tasks, delete tasks and run a task.
Press New on Task List page, and the New Task Wizard will appear. You can move the cursor by pressing the navigation
keys, press ENTER or the Select to chose a task.
After the RTD Transmitter is selected, press Enter or Select to enter the New Task General Data page.In this page, you can
input the general information of the DUT.
The Input Quantity defines which fields are required.Common fields for all quantities are: Input Quantity (Parameters), Unit
and Range.On this page, you should select an available RTD Sensor Type (as shown in Figure 10.2).
In addition, there may be some additional information: Pressure unit (for pressure instruments), decimals (flow instruments),
temperature unit (for RTDs/thermocouples) and cold junction compensation temperature (for thermocouples).
The Output Quantity includes the corresponding common fields similar with the Input Quantity. The same additional fields
are also available for the Output Quantity (as shown in Figure 10.3).
Note