If the task has been completed, the parameters cannot be edited, To use the same task for another calibration, press“Save
As”to save it as a new task and the parameters will be editable.
You can press the navigation keys Left/Right to switch the content in this page.
The Calibration Method page will appear to allow you to select between manual or automatical calibration (as shown in
Automatic Calibration
Automatic calibration is possible when the calibrator is able to generate/simulate the DUT's input signal, and measure the
DUT's output signal.
The Step Time (set point delay) setting is in use only when the Calibration Method is set to Automatic. It determines the interval
between the two calibration points.
Manual Calibration
In Manual Calibration, you have to manually set the value that the calibration point requires. Manual calibration is possible
in almost all situations, even for calibrations that can be performed automatically.
Adjust the output signal of the calibrator until the reading of the DUT displays a nominal set point, then record the reading
from the calibrator. The calibrator will calculate the DUT's theoretical reading accordingly.