53
You can save the point data and continue to the next point until all points are accepted by pressing or .
The calibrator calculates the maximum error, maximum hysteresis, zero error and span error. Depending on the value of the
maximum error, you can decide whether to adjust the DUT.
If you just make a calibration test, you can press Esc or Reject to abort the AsCal calibration.
If you do not want to save the calibration results and continue to calibrate the DUT, you can press ReCal to return to the AsCal
running page.
If you want to accept the calibration results, and save the data. Press Save then select the save options.
If the DUT is a HART-Transmitter, the Adjust will be enabled.