(2) Pressure switch calibration
Operation steps are:
① Connect the pressure switch and
.
② Press to select pressure switch
measure. (optional: please select the
triggering type in the menu for pressure
triggering measure)
③ According to sketup map, please connect
the test in terminals cables.
◎SW connect "+" of switch, red color;
◎COM connect "-" of switch, black color.
④ Follow all verification regulations, take
as a reference gauge to
calibrate (verify) the pressure switch.
ADT672
ADT672
55