8 Meadowbank Road, Carrickfergus, BT38 8YF, Northern Ireland www.advancedsensors.co.uk
Page 49 of 93
OIW-HBO-0002-EX-005
OIW EX SERIES-SIDE STREAM
OPERATION MANUAL
Rev. 004 MAY 2015
The remaining menu option is ‘Auto Tuning’, Figure 59. Auto tuning, as previously described in Section 2.3.4 — Phase
Auto Tuning, is a process that will automatically sweep through a defined frequency range to detect the optimal
operating and cleaning frequencies for the local temperature. Auto Tuning will always occur at the same point in the
measurement cycle. Auto tuning must be enabled and configured by an authorized Advanced Sensors Engineer or
Representative.
Figure 59: Auto tuning menu.
CAUTION: Adjustments to the ultrasonic system (for cleaning and homogenization) shall only be performed
by Advanced Sensors trained or authorized personnel only. Excessive application of power to
the ultrasonic system may over time result in etching the measurement window and reduce the
expected working life of the transducer.