8 Meadowbank Road, Carrickfergus, BT38 8YF, Northern Ireland www.advancedsensors.co.uk
Page 62 of 93
OIW-HBO-0002-EX-005
OIW EX SERIES-SIDE STREAM
OPERATION MANUAL
Rev. 004 MAY 2015
SEVERITY (3): LOG MESSAGE (2): ACTION:
Critical Config file read error Contact Advanced Sensors for assistance.
Critical Failed to open Us com port Contact Advanced Sensors for assistance.
Critical Us psu config mismatch Contact Advanced Sensors for assistance.
Critical Failed to set gain or offest Contact Advanced Sensors for assistance.
Critical Daughterboard not responding. Contact Advanced Sensors for assistance.
Critical Temperature sensor IO fault Contact Advanced Sensors for assistance.
Critical Failed to find spectrometer Contact Advanced Sensors for assistance.
Information ,SW BOOTED on No action required.
Information Login by user
No action required. User ID will be recorded in
the system log.
Information Config upgraded successfully No action required.
Information Successfully Loaded Config No action required.
Information Config loaded from data log file No action required.
Information User note:
No action required. User comments added to the
data log.
Information Start cycle on reboot changed No action required.
Information
Alarms table full - log file holds all alarms
raised
To clear the alarms table right click in the alarm
list and select ‘Delete All’.
Information Invalid UI configuration item Save changes and reboot the system.
Information Temperature below US cut-off
If temperature continuously falls below the US
cut-off, the US cut-off temperatures require
adjustment.
Information Oil Type Auto Switched No action required.
Information Settings Saved No action required.
Information Water Status Alarm Table Full
Clear alarms table. Right click in the alarm list
and select ‘Delete All’.
Information US Comms Active No action required.
Information System Successfully Restarted No action required.
Information New oil loaded to system No action required.
Information New oil loaded to this water line No action required.
Information UI configuration file read error
Select ‘Save Changes and Logout’. Exit and
restart the software. If the issue persists contact
Advanced Sensors for assistance.
Information Spectrometer measurement range invalid
Check parameters and amend as required. Select
to ‘Save Changes and Logout’. Exit and restart
the software. If the issue persists contact
Advanced Sensors for assistance.
Information Spectrometer control loop opened No action required.
Information Auto START Cycle No action required.
Information START Cycle No action required.