4451C EN 2013-04-30Introducing the CR 85-X
CR 85-X DIGITIZER
Equipment Classification
Class I Equipment Equipment in which protection against electric shock does
not rely on basic insulation only, but includes a power sup-
ply cord with protective earth conductor. For earth reli-
ability always plug the main power cord into an earthed
mains power outlet.
Type B Equipment Not applicable: The patient does not get in contact with
any part of the device.
Water Ingress This device does not have protection against ingress of
water.
Cleaning See section “Cleaning and Disinfection” on page 22.
Disinfection See section “Cleaning and Disinfection” on page 22.
Flammable Anes-
thetics
This device is not suitable for use in the presence of a flam-
mable anesthetic mixture with air, or in presence of a
flammable anesthetic mixture with oxygen or nitrous
oxide.
Continuous Opera-
tion
The unit is suitable to run continuously.