86
4451C EN 2013-04-30Troubleshooting
CR 85-X DIGITIZER
General procedure in case of malfunction
In exceptional situations the CR 85-X display provides comprehensive
information concerning errors and ways of correcting them. The Digitizer
status changes from ‘READY’ to one of the following:
Message
Status
indicator
Meaning Action
Warning
Red flash-
ing
Further operation is possible
without impairing the image
quality.
Follow the instructions on the
display. The warning
disappears as soon as the
problem has been solved.
Locked
Red flash-
ing
The Digitizer no longer takes
cassettes from the input
buffer. You can solve this
problem without resetting the
Digitizer.
Follow the instructions on the
display.
Error
Constant
red
This status normally requires
service or key operator
intervention.
Follow the instructions on the
display.
Caution: Never reset the Digitizer to solve a cassette or image plate jam nor
to solve communication problems with the image processing station.