43
4451C EN 2013-04-30 Basic operation (‘Operator mode’)
CR 85-X DIGITIZER
Re-erasing an image plate
At the end of a normal or emergency digitizing cycle, the CR 85-X returns an
erased image plate. However, in the following cases, you must re-erase the
image plate before re-using it in order to prevent ghost images from
interfering with the image of interest:
If the image plate has not been used for more than 3 days.
In this case, the image plate may have been exposed to scattered radiation.
If an image plate has been exposed to an exceptionally high X-ray dose.
In this case, deep layers of the image plate may still retain a latent image after
standard erasure. Leave the image plate to rest at least one day before re-erasing
it.
You can erase image plates which you have given the status ‘to be erased’ via
the ID Station or image plates which have the status ‘erased’. You can erase an
image plate or a batch of up to 9 plates.
Re-erasing image plates with status ‘erased’
To erase one or more image plates which have been erased as part of a normal
or emergency digitizing cycle:
1 Check that the CR 85-X is ready for operation:
the CR 85-X must display the operator screen with ‘Ready’ or ‘Busy’ status.
the status indicator at the top of the CR 85-X must be constant or flashing
green.
READY
CR 85-X
Status

Table of Contents

Other manuals for AGFA CR 85-X

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AGFA CR 85-X and is the answer not in the manual?

AGFA CR 85-X Specifications

General IconGeneral
BrandAGFA
ModelCR 85-X
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals