4451C EN 2013-04-30Advanced operation (‘Key-operator mode’)
CR 85-X DIGITIZER
Installing new customer parameters
Via this key-operator function you can copy new customer parameter files
(CPF-files) from a floppy disk to the hard disk of the CR 85-X and activate the
new parameters.
To install new customer parameters:
1 Check that the CR 85-X is not scanning an image plate.
If the CR 85-X is scanning an image plate, the status indicator at the top of the
machine is green and flashing.
2 In the key-operator main menu, select ‘Install’ via the Up and Down keys and
confirm.
The CR 85-X will display the Install menu.
3 In the Install menu, select ‘Configuration’ via the Up and Down keys and confirm.
Note: When an error occurs during installation, refer to Chapter 4,
‘Troubleshooting.
Note: Newer versions of the digitizer have a USB port instead of a floppy disk
drive. This manual describes how to use a floppy disk. The usage of a USB
flash drive is analogous.
Queue management
Digitizer set-up
Date and Time
Send test image
System info
Install
Save configuration
Fast preview
: ok
: quit
: select
KEY-OPERATOR
MENU
SERVICE XXXXX

Table of Contents

Other manuals for AGFA CR 85-X

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AGFA CR 85-X and is the answer not in the manual?

AGFA CR 85-X Specifications

General IconGeneral
BrandAGFA
ModelCR 85-X
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals