4451C EN 2013-04-30 Advanced operation (‘Key-operator mode’)
CR 85-X DIGITIZER
Sending test images
To check the communication between the CR 85-X and the image processing
station, you can send a test image from the Digitizer to the image processing
station.
To send a predefined test image to a specific image processing station:
1 In the key-operator main menu, select ‘Send test image’ via the Up and Down keys
and confirm.
The CR 85-X will display the Send test image menu:
2 Select the image processing station to which you want to send the test image via the
Left and Right keys.
3 Confirm your choice.
The CR 85-X will return to the operator screen.
Note: If you do not press a key within 1 minute, the CR 85-X will exit the key-
operator mode.
Queue management
Digitizer set-up
Date and Time
Send test image
System info
Install
Save configuration
Fast preview
: quit
: ok
: select
KEY-OPERATOR
MENU
SERVICE XXXXX
To process.station:
[PPname1]
: quit
: ok
: change
SEND
TEST IMAGE
SERVICE XXXXX

Table of Contents

Other manuals for AGFA CR 85-X

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AGFA CR 85-X and is the answer not in the manual?

AGFA CR 85-X Specifications

General IconGeneral
BrandAGFA
ModelCR 85-X
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals