4451C EN 2013-04-30 Equipment information sheet
CR 85-X DIGITIZER
Environmental requirements
Room temperature 20°C - 30°C
Maximum temperature change 0.5°C/min.
Relative humidity 10% - 80%
Magnetic field
Less than 5 Gauss; compliant with
EN 61000-4-8, Level 5
Sunlight exposure Not to be operated in full sunlight
Barometric pressure during operation 70 kPa to 106 kPa
Related altitude on site 3000 m to 0 m
Physical emissions
Noise emission (sound power level according to DIN 45635 Part.27)
• During scanning max. 65 dB(A)
•Standby max.46dB(A)
Heat emission (at max. throughput with 35 cm x 43 cm image plates)
•Europe 0.8 kWh
•GB 2730 BTU/hr
Cassette buffer capacity
10 cassettes of mixed sizes, both in input and output buffer
Performance
Throughput for high resolution
112 plates/h (size dependent)
Throughput for standard resolution
112 plates/h (size dependent)
End of life
Estimated product life (if regularly serviced
and maintained according to Agfa
instructions)
7 years
Preventive maintenance
Preventive maintenance frequency.
Needs to be done by an Agfa certified field
service engineer
Once per 6 months or 25000 cycles,
whatever comes first