Operation 2
3200P pH Meter User Guide 31
This window displays current time setup including year,
month, day, hour, minute, and second. To modify the time
setup, use the arrow keys to highlight the proper time cell,
then press [Setup] and input the correct value. Press [Enter]
to save. When all changes have been made, press [Enter] to
save the final setup. Press [Cancel] to exit from System Setup.
To select the display language
The meter can display either Chinese or English text. To
select the preferred language, from the initial state press
[Setup]. Choose Language Select and then select either Chinese
or English.
To set the meter to turn off automatically (Auto Power Off)
This meter can automatically turn off after a specified
period of time. You can set a time period for Auto Power Off
from 10–480 minutes. When the meter has run for the
length of time that equals to Auto Power Off setpoint, the
meter will be forced to be power off.
To disable Auto Power Off, set the time period to zero (0).
When the meter is controlled using the optional G4390A
software, Auto Power Off is disabled.
Figure 15 Set date and time