Operation 2
3200P pH Meter User Guide 37
4 Prepare one to five buffers. They can be purchased or can
be prepared by you. Place them in a constant temperature
environment to equilibrate.
5 View the pH calibration data. See “To view last calibrated
data” on page 45.
6 Press [Setup] and select the pH buffer group. See “To
select the pH Buffer Group” on page 34.
7 Return to the Setup window. Scroll to Set Config Type and
press [Enter].
8 If using standard buffers from within a group, select
Auto-recognition. Matched with the proper Select Group, the
meter will automatically recognize these buffers. If the
Figure 18 Calibration
Figure 19 Calibration setup