Chapter 6 Service and Repair
Disassembly
4
6
4 Removed the Power Supply. Disconnect the input power to the power
supply board (blue and brown wires). Disconnect the green ground
connector on the power supply board. Remove the screw securing the
power supply cover to the chassis. Slide the power supply assembly
toward the front of the instrument and remove.
WARNING Always be sure to re-attach the green ground wire to the power supply
before operating the instrument.
5 The remaining assemblies can be removed from the chassis if needed.