CALCulate:AVERage:MINimum?
Read the minimum value found during a min-max operation. The
multimeter clears the value when min-max is turned on, when power
has been off, or after a remote interface reset. [Stored in volatile memory]
CALCulate:AVERage:MAXimum?
Read the maximum value found during a min-max operation. The
multimeter clears the value when min-max is turned on, when power
has been off, or after a remote interface reset. [Stored in volatile memory]
CALCulate:AVERage:AVERage?
Read the average of all readings taken since min-max was enabled.
The multimeter clears the value when min-max is turned on, when
power has been off, or after a remote interface reset. [Stored in volatile
memory]
CALCulate:AVERage:COUNt?
Read the number of readings taken since min-max was enabled. The
multimeter clears the value when min-max is turned on, when power
has been off, or after a remote interface reset. [Stored in volatile
memory]
CALCulate:NULL:OFFSet {<
value>|MINimum|MAXimum}
Store a null value in the multimeter’s Null Register. You must turn on
the math operation before writing to the math register. You can set the
null value to any number between 0 and
±120% of the highest range,
for the present function.
MIN =120% of the highest range. MAX = 120%
of the highest range. [Stored in volatile memory]
CALCulate:NULL:OFFSet? [MINimum|MAXimum]
Query the null value.
CALCulate:DB:REFerence {<
value>|MINimum|MAXimum}
Store a relative value in the dB Relative Register. You must turn on the
math operation before writing to the math register. You can set the
relative value to any number between 0 dBm and
±200 dBm.
MIN = –200.00 dBm. MAX = 200.00 dBm. [Stored in volatile memory]
CALCulate:DB:REFerence? [MINimum|MAXimum]
Query the dB relative value.
4
Chapter 4 Remote Interface Reference
Math Operation Commands
125

Table of Contents

Other manuals for Agilent Technologies 34401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34401A and is the answer not in the manual?

Agilent Technologies 34401A Specifications

General IconGeneral
BrandAgilent Technologies
Model34401A
CategoryMultimeter
LanguageEnglish

Related product manuals