To Make Null (Relative) Measurements
Each null measurement, also called relative, is the difference between a
stored null value and the input signal.
Result = reading null value
You can make null measurements with any function except
continuity, diode, or ratio. The null operation is local to the selected
function; when you change functions, null is disabled.
To null the test lead resistance for more accurate two-wire ohms
measurements, short the ends of the test leads together and then
press
Null .
The first reading taken after you press Null is stored as the null
value in the Null Register. Any previously stored value is
replaced with the new value.
After enabling null, you can edit the stored null value by
pressing
Shift > (Menu Recall). This takes you to the
NULL VALUE” command in the MATH MENU (only if null is
enabled). Go down to the “parameter” level, and then edit the
displayed value.
The null register is cleared when you change functions, turn null off,
turn off the power, or perform a remote interface reset.
Enables null operation;
Press again to disable.
Math annunciator is on when
null operation is enabled.
To read / edit the null value, use the MATH menu.
Chapter 2 Front-Panel Menu Operation
To Make Null (Relative) Measurements
38

Table of Contents

Other manuals for Agilent Technologies 34401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34401A and is the answer not in the manual?

Agilent Technologies 34401A Specifications

General IconGeneral
BrandAgilent Technologies
Model34401A
CategoryMultimeter
LanguageEnglish

Related product manuals