To Select the Programming Language
You can select one of three languages to program the multimeter
from the selected remote interface. The language is
SCPI when
the multimeter is shipped from the factory. See alsoProgramming
Language Selection,” on page 94.
1 Turn on the front-panel menu.
A: MEAS MENU
2 Move across to the
I/O MENU choice on this level.
E: I/O MENU
3 Move down a level and then across to the LANGUAGE command.
5: LANGUAGE
4 Move down to the “parameter level to select the language.
Choose from one of the following: SCPI, Agilent 3478A, or Fluke 8840A.
SCPI
5 Save the change and turn off the menu.
The language selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
<
<