Number of Triggers
Normally, the multimeter will accept only one trigger before returning
to the “idle trigger state. You can, however, instruct the multimeter to
accept multiple triggers.
This feature is available only from the remote interface. If you set the
trigger count and then go to local (front panel), the multimeter ignores
the trigger count setting; when you return to remote, the trigger count
returns to the value you selected.
Number of triggers: 1 to 50,000. The default is 1 trigger.
The selected number of triggers is stored in volatile memory; the
multimeter sets the trigger count to 1 when power has been off or
after a remote interface reset. The CONFigure and MEASure?
commands automatically set the trigger count to 1.
Remote Interface Operation:
TRIGger:COUNt {<
value>|MINimum|MAXimum|INFinite}
Chapter 3 Features and Functions
Triggering
78