Number of Samples
Normally, the multimeter takes one reading (or sample) each time it
receives a trigger from the selected trigger source (if the multimeter is
in the wait-for-trigger state). You can, however, instruct the multimeter
to take multiple readings for each trigger received.
Number of samples: 1 to 50,000. The default is 1 sample per trigger.
The selected number of samples is stored in volatile memory; the
multimeter sets the sample count to 1 when power has been off or
after a remote interface reset. The CONFigure and MEASure?
commands automatically set the sample count to 1.
Front-Panel Operation:
3: N SAMPLES (TRIG MENU)
Remote Interface Operation:
SAMPle:COUNt {<
value>|MINimum|MAXimum}
3
Chapter 3 Features and Functions
Triggering
77