The standard event register is cleared when:
You send a *CLS (clear status) command.
You query the event register using the *ESR? (event status register)
command.
The standard event enable register is cleared when:
You turn on the power and you have previously configured the
multimeter using the *PSC 1 command.
You execute a *ESE 0 command.
The standard event enable register will not be cleared at power-on if you
have previously configured the multimeter using *PSC 0.
4
Chapter 4 Remote Interface Reference
The SCPI Status Model
141