-112 Program mnemonic too long
A command header was received which contained more than the
maximum 12 characters allowed. Example: CONFIGURATION:VOLT:DC
-113 Undefined header
A command was received that is not valid for this multimeter. You may
have misspelled the command or it may not be a valid command. If you
are using the short form of the command, remember that it may contain
up to four letters. Example: TRIGG:COUN 3
-121 Invalid character in number
An invalid character was found in the number specified for a parameter
value. Example: STAT:QUES:ENAB #B01010102
-123 Numeric overflow
A numeric parameter was found whose exponent was larger than 32,000.
Example: TRIG:COUN 1E34000
-124 Too many digits
A numeric parameter was found whose mantissa contained more than
255 digits, excluding leading zeros.
-131 Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may
have misspelled the suffix. Example: TRIG:DEL 0.5 SECS
-138 Suffix not allowed
A suffix was received following a numeric parameter which does not
accept a suffix. Example: SAMP:COUN 1 SEC (
SEC is not a valid suffix).
-148 Character data not allowed
A discrete parameter was received but a character string or a numeric
parameter was expected. Check the list of parameters to verify that you
have used a valid parameter type. Example: DISP:TEXT ON
Chapter 5 Error Messages
Execution Errors
174