Amplitude Linearity
Measurement Lower
Limit
(dB)
Upper Limit
(dB)
Measured Value
(dB)
Pass/Fail
13 dBVrms, Ch 1 -0.0615 0.061
13 dBVrms, Ch 2 -0.0615 0.061
13 dBVrms, Ch 3 -0.0615 0.061
13 dBVrms, Ch 4 -0.0615 0.061
-1 dBVrms, Ch 1 -0.105 0.104
-1 dBVrms, Ch 2 -0.105 0.104
-1 dBVrms, Ch 3 -0.105 0.104
-1 dBVrms, Ch 4 -0.105 0.104
-15 dBVrms, Ch 1 -0.33 0.318
-15 dBVrms, Ch 2 -0.33 0.318
-15 dBVrms, Ch 3 -0.33 0.318
-15 dBVrms, Ch 4 -0.33 0.318
-29 dBVrms, Ch 1 -1.551 1.316
-29 dBVrms, Ch 2 -1.551 1.316
-29 dBVrms, Ch 3 -1.551 1.316
-29 dBVrms, Ch 4 -1.551 1.316
-43 dBVrms, Ch 1 -13.823 5.088
-43 dBVrms, Ch 2 -13.823 5.088
-43 dBVrms, Ch 3 -13.823 5.088
-43 dBVrms, Ch 4 -13.823 5.088
-53 dBVrms, Ch 1 -30.116 10.896
-53 dBVrms, Ch 2 -30.116 10.896
-53 dBVrms, Ch 3 -30.116 10.896
-53 dBVrms, Ch 4 -30.116 10.896
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Verifying Specifications Agilent 35670A
Operation Verification Test Record - Four Channel
10 of 14