To set up the input resistance test
Performance Test only
This test verifies that the Agilent 35670A meets its input resistance specification.
In this test, a digital multimeter directly measures the input resistance of each
channel. The digital multimeter is set to the 1 M
range.
1 2ch
2 2ch
Agilent 35670A Verifying Specifications
To set up the input resistance test
3-37