EasyManua.ls Logo

Agilent Technologies 35670A - Cross Channel Amplitude; Cross Channel Phase

Agilent Technologies 35670A
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cross Channel Amplitude
FFT cross channel gain accuracy
Frequency response mode, same amplitude range (AC coupled,periodic chirp,uniform window,>=4Hz)
At full scale: Tested with 10 rms averages
on the -11 to +27 dBvrms ranges, and 100
rms averages on the -51 dBVrms range
±0.04 dB (0.46%)
At -20 dBfs: Tested with 200 rms averages
on the -11 to +27 dBVrms ranges, and 2000
rms averages on the -51 dBVrms range
±0.08 dB (0.92%)
Cross Channel Phase
Cross channel phase accuracy
(same conditions as cross-channel amplitude)
±0.5 degree
Agilent 35670A Specifications
Cross Channel Amplitude
1-7

Table of Contents