To load the program
For information about the program’s softkeys, see the menu descriptions starting on
page 3-49.
1 Set the Agilent 35670A Dynamic Signal Analyzer’s power switch to off ( O ), then
connect the analyzer, test instruments, and printer using GPIB cables.
2 If you have the PC Style Keyboard, option 1CL, connect the keyboard to the analyzer
using the keyboard cable (see ‘’To connect the optional keyboard’’ in chapter 2).
3 Insert the Semiautomated Performance Test Disk into the analyzer’s disk drive, then
set the power switch to on (
l ).
4 After the analyzer finishes its power-up calibration routine, press the following keys:
[
Local/GPIB ]
[
SYSTEM CONTROLLR ]
[
System Utility ]
[
MEMORY USAGE ]
[
REMOVE WATERFALL ]
[ CONFIRM REMOVE ]
[
RETURN ]
[
MORE ]
[
SERVICE TESTS ]
[ PERFRMANC TEST ]
5 Now go to one of the following procedures to continue:
• ‘’To run the program in semiautomated mode’’
• ‘’To run the program without a printer’’
• ‘’To run the program in manual mode’’
Agilent 35670A Verifying Specifications
To load the program
3-7