70 Chapter 4
Agilent 4072/4073 Preinstallation Guide, Edition 4
Receiving
Moving Cabinet, Testhead and Accessories
Moving cabinet and accessories
1. Move the items to the front of the clean room entrance. The system cabinet is equipped with casters to allow
movement of the cabinet.
CAUTION When moving the system on its casters, do not allow the casters to become caught on cables or
debris, or in cracks in the floor. The high center of gravity and the momentum caused by pushing
or pulling can cause the system to tip over, resulting in serious injury to personnel and damage to
the equipment
NOTE When moving the system, fix the mounting support under the cabinet to the uppermost position.
The collision in moving cause the system to damage.
2. Remove the outer protective vinyl cover.
3. Move the items into the buffer room.
4. Remove the inner vinyl cover.
5. Move the items to the final installation location.
CAUTION After moving the cabinet, immediately attach the cabinet using the fixing mount. If the cabinet is
not secured, the cabinet could be knocked over, the system resulting in serious injury to personnel
and damage to the equipment

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4072A Advanced and is the answer not in the manual?

Agilent Technologies 4072A Advanced Specifications

General IconGeneral
BrandAgilent Technologies
Model4072A Advanced
CategoryTest Equipment
LanguageEnglish

Related product manuals