Note
In
the
program,
the
4339B
is
referred
as \UUT".
Note
If
the adjustments
being
performed
follow
right
after
EEPROM
replacement,
change
switch
A2SW1-1
to
\CLOSE",
so
that
the
4339B
will
ignore
the
self
test
errors
at
power-on.
(Return
the
switch
to
the
\OPEN"
position
when
the
adjustments
have
been
completed.)
Refer
to
Figure
B-1
for
the
switch
location.