Figure
3-2.
Cable
Assembly
Connection
for
Interlock
Connector
Triaxial
Connector
Removal
Caution
When
accessing
the
traiaxial
connector
,
use
lint
free gloves
to
avoid
contaminating
the
connector
.
Agilent
T
echnologies
supplies
lint free
gloves
(Agilent
PN
9300-0163)
.
Removal
Procedure
1.
Remove
the
cover
while
referring
to
\Cover
Removal".
2.
Disconnect
the
two
atcable
assemblies
and
the
wire
assembly
from
the
front
side
of
the
A2
board.
3.
Remove
the
screw
which
fastens
the
cable
assembly on
the front
side of
the
A3
Ammeter
Board
Assembly
.
4.
Remove
the
screws
which
fastens
the
cable
assembly on
the left
front side
of the
chassis
.
5. Remove
the side
trim label
from both
sides of
the front
bezel.
6. Remove the four screws that secure the front panel bezel to the chassis from
both sides
.
7. Place the front panel assembly
, so that the front panel faces down.
8. Disconnect the cable assembly which includes a core from the front side of the A3
Ammeter board
assembly.
9. Remove the core
holder which holds the cable assembly disconnected in the last procedure
,
by unscrewing the screw
which secures the holder from the front side
. Unsolder the diodes
from the sleeve of the
triaxial connector
.
10. Remove the heat shrink tube from the triaxial connector.
11. Unsolder the sleeve from the triaxial connector.
12. Unsolder the coaxial cable from the triaxial connector.
13. Unscrew the nut which fastens the triaxial connector to the sub panel.
Assembly Replacement 3-15