Agilent Technologies 4339B - Check 13, Transformer Secondary Voltage

Agilent Technologies 4339B
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Check
13, Transformer
Secondary V
oltage
1.
Turn o
the
4339B
.
2.
Disconnect the
transformer
secondary
connector
from
connector
\J13"
on
the
A1
Main
Board
Assembly.
Refer
to
Figure
4-14
for
the
connector
location.
3.
Turn the
4339B on.
4.
Using
a
multimeter
,
check
if
the
transformer
secondary
voltage
is
between
13
Vand
17 V
ac.
5.
If
passed,
replace
the
A1
Main
Board
Assembly
,
otherwise
replace
the
transformer
assembly
.
Figure
4-14. Transformer
Secondary
V
oltage
Check
4-18 Troubleshooting