5 Operating in CI Mode
140 5977B Series MSD Operating Manual
To Verify NCI Performance
This procedure is for EI/PCI/NCI MSDs only.
Materials needed
OFN, 100 fg/µL 3 × 1 mL (5188-5347)
Procedure
1 Verify that the MSD performs correctly in EI mode.
2 Load the NCICH4.U tune file, and accept the temperature setpoints.
3 Select Gas A and set flow to 40%.
4 In Tune and Vacuum Control view, run CI Autotune. (See “To Perform an NCI
Autotune (Methane Reagent Gas)” on page 137.)
Note that there are no criteria for a “passing” Autotune in CI. If the
Autotune completes, it passes.
5 Run the NCI sensitivity method: OFN_NCI.M using 2 µL of 100 fg/µL OFN.
6 Verify that the system conforms to the published sensitivity specification.
Please refer to the Agilent Web site at www.agilent.com/chem for
specifications.
CAUTION
Always verify MSD performance in EI before switching to CI operation. (See “To Verify
EI System Performance” on page 108.) Always set up the CI MSD in PCI first, even if
you are going to run NCI.

Table of Contents

Other manuals for Agilent Technologies 5977 MSD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977 MSD and is the answer not in the manual?

Agilent Technologies 5977 MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977 MSD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals