7 CI Maintenance
208 5977B Series MSD Operating Manual
To Install the EI HES Radiator
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
T10 Torx screwdriver (5182-3466)
Tweezers (8710-2460)
Procedure
1 Place the radiator over the guide pins on the analyzer support and use a
T10 Torx screwdriver to secure it using the (2) M3 x 12 gold plated screws
(part number G7002-20110).
2 Use tweezers to disconnect the wires from the EI HES radiator. (See “To
Connect/Disconnect Wiring to the EI HES Radiator” on page 201.)
WARNING
The analyzers, GC/MS interface, and other components in the analyzer chamber
operate at very high temperatures. Do not touch any part until you are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the analyzer
chamber.
CAUTION
Make sure you use an antistatic wrist strap and take other antistatic precautions
before touching analyzer components.

Table of Contents

Other manuals for Agilent Technologies 5977 MSD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977 MSD and is the answer not in the manual?

Agilent Technologies 5977 MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977 MSD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals