General Maintenance 6
5977B Series MSD Operating Manual 171
To Install the EI HES
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Tweezers (8710-2460)
Procedure
1 Align the EI HES so that the slotted opening for the lens stack, where the
lens numbers are visible, is on the right side. Also, position it so the two
attachment tabs of the source mount align with the corresponding slots in
the source radiator. Slide the ion source into the source radiator until
resistance is felt.
2 Snap the source into place so that the source mount tabs are flush against
the source mount attachment surface of radiator. Some force is required to
overcome the resistance of the electrical contacts spring loaded surfaces.
3 Connect the EI HES wires. (See “To Connect/Disconnect Wiring to the EI
HES” on page 156.)
4 Install and hand-tighten the source thumbscrews. Do not overtighten the
thumbscrews.
5 Close the analyzer chamber. (See “To Close the Analyzer Chamber” on
page 196.)
6 Pumpdown the MSD. (See “To Pumpdown the MSD” on page 116.)
7 Tune the MSD. (See “To Tune the MSD in EI Mode” on page 104.)
CAUTION
Always wear clean gloves when working in the analyzer chamber to avoid
contamination.

Table of Contents

Other manuals for Agilent Technologies 5977 MSD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977 MSD and is the answer not in the manual?

Agilent Technologies 5977 MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977 MSD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals