General Maintenance 6
5977B Series MSD Operating Manual 181
3 Collect the following parts to be cleaned for an EI XTR source: (See
Figure 58 on page 182.)
Repeller
Repeller block insert
Source body
Extractor lens
Ion focus lens
Entrance lens
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.
CAUTION
If insulators are dirty, clean them with a cotton swab dampened with reagent-grade
methanol. If that does not clean the insulators, replace them. Do not abrasively or
ultrasonically clean the insulators.